Analytical Instruments

TLC-FID IATROSCAN / Analytical Instruments

IATROSCAN TLC-FID/FPD Dual Detection System

TOP Distributor

The IATROSCAN MK-6 is thin layer chromatograph measuring sample(s) by means of Flame Ionization Detector(FID) and Flame Photometric Detector(FPD). The sample developed on a thin layer quarts rod 'Chromarod' designed for IATROSCAN's exclusive use can be detected by IATROSCAN automatically and quickly. Neither troublesome coloring nor activation of thin layer for hours is required for IATROSCAN analysis. Direct detection can be made for a large variety of organic compounds, including not only those analyzable by TLC but also samples of a higher boiling point which are difficult to analyze by GC as well as those problematic to detect by LC.
IATROSCAN MK-6s with only FID is available as well.
 
  IATROSCAN TLC-FID/FPD TLC-FID Thin-Layer Chromatography with Flame Ionization Detector  


Heavy oil by IATROSCAN (TLC-FID/FPD)

Heavy oil type analysis SARA Type Analysis IATROSCAN TLC-FID Thin-Layer Chromatography with Flame Ionization Detector

Sample:Bitumem
1 Saturated H.C.
2 Aromatic
3 Resin
4 Asphaltene


Stationary phase :Chromarod-SIII

Mobile phase :

1st. Hexane 100% 10cm
2nd. Toluene 100% 6cm
3rd. Dichloromethane:Methanol 57:3 2.5cm


Lipids by IATROSCAN (TLC-FID/FPD)

Lipids analysis IATROSCAN TLC-FID Thin-Layer Chromatography with Flame Ionization Detector

Sample:Standard mixture
1.Cholesterol ester
2.Triglyceride
3.Cholesterol
4.Phosphatidyl ethanolamine
5.Phosphatidyl choline
6.Phosphatidyl inositol
7.Sphingomyelin
8.Lysophosphatidyl choline


Stationary phase :Chromarod-SIII

Mobile phase :
1st. Chloroform:Methanol:Water: 25%Ammonia
47:20:2.5:0.28 7cm
2nd.Hexane :Diethyl ether 63:7 10cm


Glycerides by IATROSCAN (TLC-FID)

Glycerides analysis IATROSCAN TLC-FID Thin-Layer Chromatography with Flame Ionization Detector

Sample:Standard mixture

1.TG (tripalmitin)
2.FA (palmitic acid)
3.1,3-DG (1,3-dimyristin)
4.1,2-DG (1,2-dipalmitin)
5.MG (monopalmitin)


Stationary phase : CHROMAROD-SIII


Mobile phase :
Benzene :Chloroform : Acetic acid
50 : 20 : 0.7 10cm

pageTop


Diagnostics Testing | Nonclinical & Clinical Services | R&D | Food Sanitation & Hygiene Analysis | Doping Tests | Corporate Profile | Quality & Environmental Policy | Recruiting Info